Article Text

Download PDFPDF
Extended report
Ability of ultrasound imaging to detect erosions in a bone phantom model
  1. Juhani M Koski1,
  2. Eeva Alasaarela2,
  3. Irma Soini3,
  4. Kari Kemppainen3,
  5. Ullamari Hakulinen4,
  6. Jari O Heikkinen4,
  7. Mikko S Laasanen5,
  8. Simo Saarakkala6,7
  1. 1Department of Internal Medicine, Mikkeli Central Hospital, Mikkeli, Finland
  2. 2Division of Rheumatology, Department of Internal Medicine, Oulu University Hospital, Oulu, Finland
  3. 3Department of Radiology, Rheumatism Foundation Hospital, Heinola, Finland
  4. 4Department of Clinical Physiology and Nuclear Medicine, Mikkeli Central Hospital, Mikkeli, Finland
  5. 5Information Technology R&D Unit, Engineering Kuopio, Savonia Polytechnic, Kuopio, Finland
  6. 6Department of Physics and Mathematics, University of Eastern Finland, Kuopio, Finland
  7. 7Department of Health Technology and Informatics, Hong Kong Polytechnic University, Hong Kong, China
  1. Correspondence to Dr Juhani M Koski, Department of Internal Medicine, Mikkeli Central Hospital, Etelä-Savo Hospital District, Porrassalmenkatu 35-37, 50100 Mikkeli, Finland; f.koski{at}fimnet.fi

Abstract

Objectives The authors examined the validity, interobserver reliability and interscanner variation in detecting bone erosions with ultrasonography using a custom-made phantom.

Methods 21 bovine bones were used. Artificial erosions were made into 15 bones and six bones were left as controls. In the processed bones the numbers of erosions, their depths and widths varied between 1–7, 1–4 and 1.5–5 mm, respectively. Each bone was coated with polyvinyl alcohol cryogel to mimic overlying soft tissue and to hide the erosions. Four musculoskeletal sonography experts scanned the 21 blind-coded phantoms using one of the three sets of ultrasound equipment. Finally, quality assurance measurements of the ultrasound equipment was carried out using two additional bone samples.

Results The sonographers detected the erosions successfully with ultrasound. The mean correlation coefficient for a correct result in terms of the number of erosions detected was 0.88 (range 0.75–0.975). The overall Cohen's κ coefficient for interobserver agreement was 0.683 in terms of discrimination between healthy bones and bones with erosions. The different sets of equipment showed that their overall performance was equal.

Conclusions The sonographers had good correlations with the number of erosions and they were successful in separating healthy bones from bones with erosions. It seems that neither depth nor width is crucial but that in experimental conditions a 1.5 mm erosion width was the limit for the resolution with current ultrasound equipment. Ultrasound is a valid and reliable method of detecting cortical bone erosions in vitro, when the round erosion is at least 1 mm deep and 1.5 mm wide.

Statistics from Altmetric.com

Footnotes

  • Funding Financial support was received from the Academy of Finland (project 127198, Kuopio, Finland) and the Ministry of Education, Finland, to the University of Eastern Finland (project 5741), Kuopio University Hospital (EVO grant).

  • Competing interests None.

  • Provenance and peer review Not commissioned; externally peer reviewed.